EQUIPMENT / CO2 INCUBATORS

Monitoring of CO2 incubators

Temperature, humidity and CO2 are measured with one sensor inside the chamber. Three parameters in the same graph, so a deviation can be seen in context rather than in isolation.

CO2 incubator with sensor through the access port

How do you monitor a CO2 incubator?

With a triple sensor placed inside the chamber that measures temperature, relative humidity and CO2 concentration at the same time. The sensor is small enough for compact incubators and connects to the data logger outside with a flat cable through the access port, so the door still closes tightly.

The sensor is powered by the data logger and needs no external power inside the chamber. All three parameters are logged in the same graph, so you can see whether a drop in CO2 coincides with repeated door openings.

Limits and delays are set according to your own SOPs and what is being cultured.

The incubator measures itself

The display on an incubator shows what the unit itself believes. It is the unit's own sensor reporting on the unit's own condition, and the data is rarely stored in a way you can present.

It only becomes a problem when the unit's sensor drifts. The display then still shows the right number, while the conditions inside the chamber are different. And a CO2 level that slips slowly because of a leaking door or a low cylinder does not look like anything on a display. It only shows up in a graph over the day.

An independent sensor with traceable calibration measures what the cultures actually experience, and stores it.

IVF or cell laboratory with several incubators

What we monitor in the incubator

Temperature, humidity and CO2

One sensor for all three parameters, in the same graph. No extra power inside the chamber.

Temperature, humidity and CO2 Smart-Sensor

The door

A dry contact shows whether repeated door openings are causing a drop in CO2 or humidity.

Dry contact cable

CO2 supply

The cylinder pressure or the supply system's alarm output via dry contact or 4-20 mA.

Universal Smart-Sensor

Room conditions

Temperature and humidity in the laboratory itself, which affect the chamber every time it is opened.

Temperature and humidity sensor

Escalation

Call, text message and email to the person on call, then on to the next, until someone acknowledges.

OCEAView

Calibration

Traceable certificate on the sensor, and recalibration by exchange without interrupting the culture.

Accredited calibration

OCEAView from DICKSON, set up and looked after by us

The hardware is DICKSON's. The software is DICKSON's. Everything between the equipment and your day-to-day work is ours. And once the system is in place, it is yours. You buy it, you own it, and no lease agreement decides when you can switch, expand or stop. We handle the maintenance so it keeps running.

OCEAView

The whole setup on one screen. Watch mode gives a combined status overview of the equipment you have access to, with up to 100 units per page, and every graph can be zoomed down to the individual minute. Plan View also places the points on your own floor plan, so the person on call can see which room it concerns.

OCEAView Watch mode with a status overview of all measuring points
View OCEAView →

An audit trail that cannot be altered

Every measurement, alarm and acknowledgement with timestamp and user. That is what an auditor asks for.

Reports in one click

Choose a period and export the report as PDF or CSV. Ready before anyone asks.

Cloud or your own servers

On-premises runs on your site with LDAP, and you control updates yourselves, so the qualification holds.

DICKSON

The manufacturer

Builds data loggers, sensors and OCEAView. French and American production, its own accredited calibration laboratories and CE, FCC and IC approved equipment.

Djurhuus IT

Distributor since 2018

We advise on placement, design the setup, install, set up alarms and on-call rotas, train your staff in the system and manage calibration afterwards.

You

Owner of the system

The system is yours. You set limits, response times and requirements according to your own SOPs. We supply the system and the documentation, handle validation, on our own or in close collaboration with you, and maintain it afterwards.

Frequently asked questions

Is the sensor small enough for our incubator?

Yes. The triple sensor is designed for compact incubators, and the flat cable is run through the access port, so the door still closes tightly.

Does the sensor need power inside the chamber?

No. It is powered from the data logger outside the incubator, so there are no extra wires or adapters in the chamber.

Why not just use the incubator's own alarm system?

Because it measures with the unit's own sensor and rarely stores data in a way you can document from. An independent measurement with traceable calibration is something other than the unit's own display.

How do we avoid an alarm at every door opening?

Each limit has its own delay. The value must stay outside the limit for the number of minutes you decide before the alarm goes off, and a dry contact cable shows at the same time whether the door was open.

Can we see temperature, humidity and CO2 together?

Yes. The three parameters are shown in the same graph, so a drop in CO2 can be seen in the context of humidity and temperature instead of in isolation.

How often does the sensor need calibrating?

You define the interval yourselves based on your SOPs, typically every 12 months. The system notifies you when a sensor's calibration is about to expire, and you choose the notice period yourselves. The certificate is retrieved directly from the sensor.

De bruger os allerede

TMJRegion SyddanmarkRegion HovedstadenRegion SjællandSDUDTUFertin PharmaTransport Billund

Kom hurtigt i kontakt med den rette specialist.

Vi hjælper med alt fra nye monitoreringsløsninger til kalibrering, validering og teknisk support.

Generelle henvendelser

Nye projekter

IT & teknik

Administration

Besøg os

Måløv Byvej 229
2760 Måløv

Ring til os

+45 72 22 02 05
Man-fre, 08-16

Fortæl os, hvad du har brug for.

Thank you! Your submission has been received!
Oops! Something went wrong while submitting the form.

Vil du se, hvordan det ser ud hos jer?

Book en gennemgang. Vi kommer forbi, ser på udstyr og lokaler og rådgiver om, hvor følerne skal sidde. Derefter får du et skræddersyet forslag. Uden forpligtelser.

Book en gennemgang